Omega Engineering OMB-NETSCAN 1500 User Manual page 176

Ethernet/internet-based data logging & control instrument
Table of Contents

Advertisement

API Command Reference
TYPE
EXECUTION
SYNTAX
DESCRIPTION
The Program Trigger Times (
acquisition that is configured for start and/or stop on Absolute Time as configured by the Set Trigger
Configuration (
The Program Trigger Times command is to be used in concert with the Set Trigger Configuration command when
it is desired to begin and/or stop an acquisition on certain times and/or dates. The values of the start and stop
times are only used when the start and/or stop events have been configured for Absolute Time.
If only the start time is needed, the
only the stop time is needed, the
In addition to the ability to start/stop on a particular time/date the unit has the ability to start/stop at a particular
time, independent of the date. If the date for the start/stop time is of no concern, then the date portion of the
time/date field should be set to
time regardless of the date. This feature is particularly useful when using the auto-rearm feature of the unit since
it will allow the collection of data at a particular time each day without user intervention.
EXAMPLE
PRINT#1,"S12:54:00.0, 01/01/93X"
PRINT#1,"P01:00:00, 0, 01/01/93,
02:00:00.0, 01/01/93X"
PRINT#1,"T11,11,0,0X"
A-36
P
- Program Trigger Times
Channel
Deferred
Pstart,
Set the trigger start and/or stop times for an acquisition configured for start
stop
and/or stop on absolute time, where
time/date stamp format:
MM
is minutes,
is day, and
dd
P?
Query current start and stop times in the form:
P
) command may be used to set the trigger start and/or stop times for an
T
) command.
stop
argument should be set to
start
argument should be set to
00/00/00
HH:MM:SS:T,mm/dd/yy
SS
is seconds,
T
is tenths of a second,
is year.
yy
00:00:00.0,00/00/00
00:00:00.0,00/00/00
. In this way, the unit can start or stop an acquisition at a particular
' Set the internal clock of the unit
' Set start (1:00 am) and stop (2:00 am) acquisition
times
' Configure for start/stop on Absolute Time
Appendix A
start
and
stop
are in the standard
where
H
is hours,
mm
is month,
Pstart,stop
. Likewise, if
.
NetScan User's Manual

Advertisement

Table of Contents
loading

This manual is also suitable for: