Safety Inspection And Tests After Repair And Maintenance; General Directives; Safety Components; Checking The Protective Earth Connection - Philips PM 5390 Service Manual

Rf
Table of Contents

Advertisement

SAFETY
INSPECTION
AND
TESTS
AFTER
REPAIR
AND
MAINTENANCE
IN
THE PRIMARY
CIRCUIT
GENERAL
DIRECTIVES
Take
care that
creepage
distances
and
clearances have
not been reduced
Before
soldering, wires:
should be bent through
the holes of solder
tags,
or
wrapped
round
the
tag
in
the
form
of an
open
U,
or,
wiring
ridigity shall
be maintained by
cable
clamps
or cable
lacing.
^
Replace
ail
insulating
guards
and
-plates.
SAFETY COMPONENTS
Components
in
the
primary
circuit
may
only be renewed
by components
selected
by
Philips,
see
also
chapter
9.1
.
CHECKING THE
PROTECTIVE
EARTH CONNECTION
The
correct
connection and condition
is
checked
by
visual
control
and
by
measuring
the
resistance
between
the
protective
-lead
connection
at
the
plug
and
the cabinet
/fra
me.
The
resistance
shall
not be
more
than
0.5
r2.
During
measurement
the
mains
cable
should be
moved.
Resistance variations
Indicate
a
defect.
CHECKING THE
INSULATION RESISTANCE
Measure
the
insulation resistance
at
U
=
500
Vdc
between
the
mains connections and
the protective
lead
connections. For
this
purpose
set
the
mains
switch to
ON. The
insulation resistance
shall
not be
less
than 2
M^2.
Note;
2
is
a
minimum
requirement
at
40^ C
and
95
%
relative
humidity.
Under
normal
conditions the
insulation
resistance
should be
much
higher (10 to
20
MO).