Philips 150S4FG/00 Service Manual page 37

15" tft lcd colour monitor
Hide thumbs Also See for 150S4FG/00:
Table of Contents

Advertisement

ZERO CALIBRATION
Zero calibration is performed to determine the output of the measuring probe when no light reaches the sensor and to set this as the zero
point to which all other measurements are referenced. Zero calibration must be performed after the POWER switch has been set ON
brfore taking any measurements.
To perform zero calibration :
* Before performing zero calibration, check that the measuring probe has been connected to probe
connector P1.
1. Check that the POWER switch is set to ON.
2. Set the measuring/viewing selector to the
(viewing) position. (An image can be seen in the
viewfinder, but no light will reach the sensor.)
3. Press 0-CAL.
* If zero calibration is being performed immediately
after the POWER switch has been set to ON, press
0-CAL after "PUSH 0-CAL KEY" appears in the
liquid crystal display.
4. Set the measuring/viewing selector to the
Measurements will be started immediately.
1. POWER switch
3. 0-CAL
Colour adjustment
DARKEN PROBE
DARKEN PROBE
PUSH 0-CAL KEY
ZERO CALIBRATION
CH00
E1 [
position.
150S4 LCD
Go to cover page
Display after POWER
switch is set to ON
Approximately 5 seconds
Press 0-CAL,
Performing zero
calibration
P1
Zero calibration
]
completed
Measuring/viewing selector
Index
37

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents