Apply prepared calibration weight*
just in the centre of the weighing
pan. The display will start blink-
ing to adjust the span exactly.
When the calibration is completed,
the display will return to the
weighing mode.
Remarks
1. During Operation 2, if Fkey is pressed first, the mode will return to
weighing.
2. The calibration is available with 1/2 of the scale capacity. Nevertheless,
a calibration weight closer to F.S. is recommended to use for accurate
calibration.
3. It is recommended to use a calibration weight of better accuracy than
the division of the scale.
4. Problems in calibration performance will be displayed by any of following
error messages, which disturbs the calibration. Check weight.
o
. £
r r
: The calibration weight is over the full capacity.
I - E r r • The calibration weight is less than 1/2 of the capacity.
c
1
- E r r : The data error exceeds 0.4%. Or perhaps the scale may
be defective. Contact the shipper.