Specifications - Velleman DVM6243 Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
3.3 Inductance (L) measurement procedure
1.
Select the range switch for the maximum expected inductance.
2.
Connect the alligator clips to the inductor leads or insert leads of the inductor into meter's measuring socket.
3.
Read the display. The measuring value is direct reading and the electrical unit (mH, H) is indicated.
4.
When only the figure "1" is displayed, it indicates over range situation and the higher range has to be selected.
5.
If the display indicates one or more zeros, shift to the next lower range scale to improve the resolution of the
measurement.
3.4 Capacitance (C) measuring procedure
1.
Select the range switch for the maximum expected capacitance.
2.
Connect the alligator clips to capacitor leads or insert leads of the capacitor into meter's measuring socket.
3.
Read the display. The measuring value is direct reading and the electrical unit (nF, µF) is indicated.
4.
When only the figure "1" is displayed, it indicates over range situation and the higher range has to be selected.
5.
If the display indicates one of more reading zeros, shift to the next lower range scale to improve the resolution of
the measurement.

4. Specifications

4.1 Features
21mm LCD display provided. MAX. indication 1999.
6 ranges from 2nF to 200µF
4 ranges from 2mH to 2 H
High accuracy in measuring
Zero automatic
Dual - Slope integration A/D converter
Overload indication of "1"
Safety designed test probe
Size : 31.5 mm x 91 mm x 189 mm (H x W x L)
Weight : 300g (including battery).
4.2 Specifications
Accuracy is specified for a period of one year after calibration and at 18°C to 28°C (64°F to 82°F) with relative
humidity to 80%.
4.3 General
POWER SUPPLY
LOW BATTERY INDICATION
FUSE PROTECTION
ZERO ADJUSTMENT
OPERATING TEMPERATURE
STORAGE TEMPERATURE
DVM6243
9V battery NEDA 1604 or 6F22 006P
"
" appears on the display
100mA/250V
Automatic (except 2nF range, typically have 0-3pf circuit stray
capacitance)
0°C to 40°C (32°F to 104°F)
-10°C to 50°C (14°F to 122°F)
3