USER GUIDE
SAFETY • INSTALLATION & INTEGRATION • OPERATING INSTRUCTIONS • MAINTENANCE • SERVICE
Outputs
Up
CONDENSER FAN=0%
Down
Outputs is used to monitor the state of DC outputs.
ID
Description
Evap Fan
Evaporator Fan
Condenser Fan
Condenser Fan
Light
Zone Lighting
To access Outputs
1. Press
to select "Outputs".
2. Press
.
3. Press
and
to scroll through available
information.
To exit the Input Status menu, press
to Menu" and press
to confirm.
Offsets
Up
Can be displayed in Celsius
Down
NOTICE
Customer care MUST be notified and approve of
any changes to the differential before they are
made. Failure to notify customer care will void
the warranty.
Select
RETURN TO MENU
OUTPUTS
EVAP FAN=0%
States
0 - 100%
0 - 100%
Off - Low - Med - High
to select "Return
Select
RETURN TO MENU
OFFSETS
ZONE = 1°F
EVAP = 1°F
Offsets are used to adjust or correct thermistor readings.
Offset values are added to the current thermistor reading
and are then used by the control board to determine
cooling and defrost cycle times. Offsets have a range of
+/- 10°C. Corrected values may be viewed through the
customer "All Temps" menu or TTY output.
To access Offsets
1. Press
to select "Offsets".
2. Press
.
3. Press
and
to scroll through available
thermistors.
To change offset
4. Press
, the selected thermistor will begin to flash.
5. Press
or
to modify offset value.
6. Press
to confirm setting.
To exit the Offset menu, press
Menu" and press
to confirm.
Self Test
Up
Down
u-line.com
to select "Return to
Select
RETURN TO MENU
SELF TEST
NO ERRORS
try "RELAY TOGGLE"
Control Operation - Service 5