Operating Instructions

1. Push the cannula breathing tube firmly on to the
oxygen tube connector.
2. Adjust the cannula breathing tube to the proper
position so that you will be able to breath the oxygen
comfortably.
3. Turn the flow rate selection knob on top of the Spirit
clockwise until the prescribed flow rate (numeral) is
visible in the knob "window" and a positive detent is
felt.
4. With each breath, you should now be receiving a set
dosage of oxygen gas. The LED will illuminate green
with each breath, signaling that the Spirit
system is being triggered. When the LED illuminates
red with each breath, it is time to replace the
batteries.
Note: Make sure that the flow selector knob fully
engages at the selected setting. If the knob is
between settings, the unit will turn off and then back
on and deliver a prolonged burst of O2 as it powers
up.
Note: The Spirit
maximum breathing rate of 40 BPM. With breathing
rates greater than 40 BPM oxygen dosage may not
be delivered with each breath.
®
Note: When turning the flow control knob from
either (0) or (CF) to a flow setting, there will be a
prolonged burst of O2 delivered through the
cannula*. Please do not be alarmed as this is
normal operation.
®
delivery
12
*Cannula is not included.
Spirit
®
®
delivery system is designed for a
ortable