29Adc (6033A).
10Adc (6038A).
Check that the unit's CC LED remains lighted.
d.
Check that the rms noise current measured by the current probe and rms voltmeter is no more than:
15mA rms (6033A).
5mA rms (6038A).
Initialization Procedure
Follow the procedure if either the HP-IB assembly has been replaced, or the EEPROM (U70) has been replaced:
1.
Install the HP-IB assembly in the unit.
2.
Turn the power on and depending on your unit's model number, send string:
"EEINIT 6033"
or
"EEINIT 6038''.
3.
Turn the power off, wait 5 seconds, then turn the power back on.
4.
If the HP-IB assembly has been replaced, calibrate the unit.
28
Figure 2-9. CC PARD Test Setup