Buzzer - Miele G 7893 Operating And Installation Instructions

Miele laboratory glassware washer operating and installation instructions
Table of Contents

Advertisement

Buzzer

The buzzer can be programmed for the following functions:
– a continuous tone at the end of a program,
– an error signal that beeps in 1-second intervals, and
– a sample-collection signal (allows a sample of the final rinse water to be
collected) that beeps in sets of 3 followed by a 2-sec pause. Contact Miele
Technical Service to program the sample-collection
signal.
The numbers shown in the display correspond to the following settings:
20
Buzzer off
21
End of program (factory setting)
22
Error
23
End of program + Error
24
Sample-collection signal
25
End of program + Sample-collection signal
26
Error + Sample-collection signal
27
All functions (End of program + Error + Sample-collection signal)
Step
^ Press the "Drying" 0 button until E04 appears in the
display.
^ Turn the program selector to "Inorganica".
^ Press and hold the "Display" ß button until the
required value appears in the display, e.g., 26.
Value selected:
The buzzer will sound for 30 sec during the selected setting.
To switch it off earlier:
– Turn the program selector to f.
– Press the "Display" ß button.
– Interrupt the power supply, e.g., open the door.
Programming functions
Display / Result
E04
21
26
37