Specifications - Bioland 2005 Instruction Manual

Upper arm automatic blood pressure monitor
Table of Contents

Advertisement

SPECIFICATIONS

Measuring Method:
Oscillometric System
Indication:
Digital LCD Display
Measuring Range:
Pressure: 0~300mmHg, Pulse: 40~200 Beat/min.
Accuracy:
Pressure ±3mmHg; Pulse ±5%
Inflation
Electric Pump Inflation
Deflation:
Pressure Release Valve
Memory:
128 Memories
Power Source:
1.5V "AA" x 4 Alkaline Batteries
Battery Life:
Approx. 250 Times (Once a Day, Room Temperature 22
Working temp. & Humidity:
40~104
Storage Temp. & Humidity:
4~140
O
Outside Dimensions:
112mm(W) x 140mm(D) x 70mm(H)
Weight:
Approx. 534g (Apparatus and Battery)
Accessories:
Cuff (fit arm size 22~32cm) , Instruction Manual
Operation Voltage
6 V DC
* Specifications may be changed without notice in the event of improvement being made.
F (5~40
C), 30~85% RH
O
O
F (-20~60
C), 10~95% RH
O
-13-
If you have trouble in using the unit, please check the following points first.
ERROR DISPLAY
Nothing is displayed
when you push the
POWER button or
battery icon flash
E1: Can't normally
increase pressure
E2, E7,E8,E9:
Measurement failure
C)
O
E3. Inflate pressure too high
E4: Have shaking while
measurement
E5: Automatically pressure
release too fast
E6: Automatically pressure
release too slow
Battery icon on
The systolic pressure
value or diastolic
pressure value too high
The systolic pressure
value or diastolic
pressure value too low
Note: Relax for at least 10 minutes and then take another measurement.
TROUBLESHOOTING
POSSIBLE CAUSE
HOW TO CORRECT
No battery installation
Insert batteries
Battery worn out
Replace new batteries
The polarities of batteries
Insert battery in the correct
placed wrongly
polarities
Check your cuff if any air
Replace cuff with new one
leakage
Arm cuff not properly adjust
Renew adjust cuff position and
or placement, hand or body
tightness, and keep hand still
shaking
Re-measurement or send back
dealer for re-calibrate pressure
Hand or body shaking while
Keeping static and correct gesture
measurement
to measure again
Cuff too tight
Readjust cuff and measure again
Cuff too loose
Battery low power
Replace battery, and measure again
1. The cuff was held lower
than your heart
2. The cuff was not
attached properly
3. You moved your body or
Keeping correct position and
spoke during measurement
gesture to measure again
1. The cuff was held higher
than your heart
2. You moved your body or
spoke during measurement.
-14-
611-200500-000
VER.1.1, 11/2006