Specifications; Troubleshooting - Bioland 3002 Instruction Manual

Portable wrist blood pressure monitor
Table of Contents

Advertisement

SPECIFICATIONS

Oscillometric System
Measuring Method:
Digital LCD Display
Indication:
Measuring Range:
Pressure: 0~300 mmHg;
Accuracy:
Pressure: ±3 mmHg; Pulse ±5%
Electric Pump Inflation
Inflation
Pressure Release Valve
Deflation:
128 Memories
Memory:
1.5V "AAA" x 2 Alkaline Battery
Power Source:
Approx. 250 Times (Once a Day, Room
Battery Life:
Temperature 22
40~104
Working Temp. & Humidity:
Storage Temp. & Humidity:
4~140
67mm(W) x 76mm(D) x 32mm(H)
Outside Dimensions:
Approx. 128g (Apparatus and Battery)
Weight:
Wrist Cuff (Fit size 13.5-19.5cm),
Accessories:
Operation Voltage
*Specitications may be changed Without notice in the event of
lmprovement being made.
Pulse: 40~200 Beat/min.
o
C)
o
o
F (5~40
C), 30~85% RH
o
o
F (-20~60
C), 10~95% RH
Instruction Manual
3V DC
-21-

TROUBLESHOOTING

If you have trouble while using the unit, please check the
following points first.
ERROR DISPLAY
POSSIBLE CAUSE
Nothing is displayed
No battery installation
when you push the
Battery worn out
POWER button or
The polarities of batteries
batter yicon flash
placed wrongly
E1: Can't normally
Check your cuff if any air
increase pressure
leakage
E2, E7, E8, E9
Wrist cuff not properly
Measurement failure
adjust or placement,
hand or body shaking.
E3. Inflate pressure
too high
E4: Have shaking while
Hand or body shaking while
measurement
measurement
E5: Automatically pressure
Cuff too tight
release too fast
E6: Automatically pressure
Cuff too loose
release too slow
"
" Battery icon on
Battery low power
The systolic pressure
1. The hand with the cuff was
held lower than your heart
value or diastolic
pressure value too high
2. The cuff was not
attached properly
3. You moved your body or
spoke during measurement
1. The hand with the cuff was
The systolic pressure
held higher than your heart
value or diastolic
2. You moved your body or
pressure value too low
spoke during measurement.
Note: 1) Relax for at least 10 minutes and then take another
measurement.
HOW TO CORRECT
Insert batteries
Replace new batteries
Insert battery in the correct
polarities
Replace wrist cuff with new one
Renew adjust wrist cuff position
and tightness, and keep hand still
Re-measurement or send back
dealer for re-calibrate pressure
Keeping static and correct gesture
to measure again
Readjust cuff and measure again
Replace battery, and measure again
Keeping correct position and
gesture to measure again
-22-