To turn the ventilator on, press the power switch at the rear of the device. Wait until the Treatment screen
is displayed.
To turn the ventilator off, press the power switch at the rear of the device. Follow the onscreen
instructions.
Accessing Clinical mode

Treatment settings

Selecting the mask type and performing a Learn circuit enables to calculate the impedance and leak for
controlled therapy pressure.
Learn Circuit configuration
Do not connect patient interfaces prior to performing the Learn Circuit. Patient interfaces include any
components placed after the ResMed Leak Valve (eg, HMEF , catheter mount, tracheostomy tube).
With an external humidifier
Antibacterial filter
ResMed Leak
Valve
Air
tubing
External Humidifier
Press
and
until the duration options appear. Follow the onscreen
instructions.
To return to the Patient menu, press
simultaneously for at least three seconds.
Program
Ramp time
Warm-up humidifier
Yellow arrows display at the top or bottom of the pressure
bar graph will show if a machine trigger or cycle occured. If
the inspiratory phase extends beyond the spontaneous cycle
by Ti Min, a red arrow is displayed.
Without an external humidifier
4
simultaneously for three seconds
Antibacterial filter
Air tubing
and
ResMed Leak
Valve