Chapter 3
13
Adjust the field diaphragm.
with the field diaphragm.)
Turn the field diaphragm ring until the diaphragm image just circumscribes the viewfield.
14
Perform the microscopy. ⇒P18
Whenever the objective is changed, adjust the condenser aperture and field aperture
diaphragms.
15
Turn off the power switch.
Press the power switch to the "○" position.
Microscopy Procedure (Quick Reference)
(This adjustment is only for microscopes
7
3