Set Of Date, Time, Log Interval (Tools/Date Time) - Panasonic MDF-U700VX-PB Operating Instructions Manual

Ultra-low temperature freezer
Table of Contents

Advertisement

VARIOUS SETTING (MENU/Tools)

Set of date, time, log interval (Tools/Date Time)

A date, time and a Log interval configuration screen (Date Time) are chosen on the top screen (Select
Tools). Then, each parameter is set up on the date, time setup screen (Date Time).
1
2
3
4
5
Each parameter setting range:
1. Date: Present date is indicated. (Example : in case of December 1, 2010) (Refer to a table 10.)
2. Time: Present time is indicated. (Example : in case of 12:00:00) (Refer to a table 10.)
3. Door Delay: The variable range of the door alarm delay is 0 to15 minutes. (Factory setting is 2
minutes.) (Page 29 Refer to a table 11)
4. Log Interval: Settable between 2 minutes and 30 minutes. (Factory setting is 15 minutes.) With 15
minutes interval, recording for about 5 weeks is available.
5. Comp Delay: Delay time for high temperature side/low temperature side compressor after power
failure. The settable range is between 3 minutes until 15 minutes. (Factory setting is 3 minutes.)
Table 10 Setting of date and time (When time is set up at 12:00:00 on December 1, 2010 )
Description of operation
1
Press the MENU button.
2
Press the shift key, highlight the
"Tools" menu window, and press the
3
ENTER key.
4
Select Date by pressing Shift keys..
Figure input key is pressed in order
5
of 1,0,1,2,0 , 1.
6
Select Time by pressing Shift keys.
Figure input key is pressed in order
7
of 1, 2, 0, 0, 0, 0.
8
Press the ENTER key.
9
Press the MENU button.
10
Press the ENTER key.
Key operated
Top screen is displayed
-----
The menu window is indicated. And the
MENU
"set" is highlighted.
Select Tools window is indicated.
ENTER
The Date
1
0
1
2
The numerical value of the date is
changed to 10/12/01.
0
1
Time is highlights.
1
2
0
0
The numerical value of the time (Time) is
changed at 12: 00:00.
0
0
ENTER
Door Delay is highlighted.
The menu window is indicated. And the
MENU
"OK" is highlighted.
Date time is memorized, and the top
ENTER
screen is indicated.
28
Indication after operation
is highlighted.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mdf-u700vx-pkMdf-u700vx-peMdf-u700vxc-pa