FREEZER COMPONENTS
1. Outer door: To open the door, grip the handle. On closing, lock the door latch completely.
2. Inner door: The operation of the inner door should be quick to minimize the temperature rise in
chamber.
Lock the inner door latch completely when the inner door is closed.
The inner door is
removable for cleaning or defrosting. See page 36 "Routine maintenance".
3. Inner door latch: Always lock the inner door latch when the inner door is closed.
4. Control panel (on the upper front of the outer door): Used for temperature setting and indication
of operating status is displayed on the panel. See page 10 for details.
5. Door latch: Always lock the latch when the outer door is closed.
6. Magnetic door gasket: This provides a tight door seal and prevents cold air leak. Keep clean.
7. Access port (rear and bottom):
This is used for leading a cable and sensor of a measuring
equipment, or nozzle of backup cooling kit to chamber.
8. Air intake vent (grille): Do not block this vent to keep the proper cooling performance.
9. Leveling foot: The height of the freezer can be adjusted by this screw type foot. Keep the unit in
level at the installation.
10. Caster: 4 casters are provided to facilitate moving of the cabinet. For the installation, adjust the
leveling feet so that the front 2 casters cannot contact with the floor.
11. Space for temperature recorder: A temperature recorder (optional component) can be attached
here. See page 45 "Temperature recorder (Option)".
o
12. Lock: Turn clockwise to 180
with a key and the outer door is securely locked.
13. Air intake port: When a door is closed and opened soon, this air intake port is used. See page 11
14. Fixture (on back side): 2 fixtures are provided as spacers between the cabinet and wall and also
serve as hooks to fix the unit. See page 13 "Installation".
15. Power switch: This is for turning ON/OFF the power to the unit.
16. Remote alarm terminal: This is used to notice an alarm condition of the unit to remote location.
Refer to page 11 "Remote alarm terminal".
17. Battery switch: This is a switch for a battery for power failure alarm. Normally, turn on this switch.
Be sure to turn off this switch if the freezer is not in operating for the long period (over one month).
WARNING
Fix the shelf stoppers and shelves securely. Incomplete installation may cause injury or damage.
Never touch the storage items with wet hands. Touching with the wet hands may cause frostbite
9