GUIDANCE AND MANUFACTURER'S DECLARATION
(Table 2)
Guidance and manufacturer's declaration – electromagnetic immunity
OMRON BPM is intended for use in the electromagnetic environment
specified below. The customer or the user of this OMRON BPM should
assure that it is used in such environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601
Compliance
Test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for
power
supply lines
Not
Applicable.
±1 kV for
input/output
lines
±1 kV
line(s) to
line(s)
Not
Applicable.
±2 kV
line(s) to
earth
Electromagnetic environment
– guidance
Floor should be wood, concrete,
or ceramic tile. If f loors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Not Applicable.
Not Applicable.