SECTION 20—4-WAY VALVE/MANIFOLD ASSEMBLY
Tie
Wrap
Manifold
Vinyl Tube
Exhaust
"F" Tube
Tube
Clamp
Tie Wrap
Tie
Clamp
Wrap
4 Way Valve/
Manifold Assembly
Vinyl Tube
DETAIL "B"
DETAIL "A"
Manifold Assembly
4-Way Valve
4-Way Valve/
Manifold
Assembly
2
4
1
3
NOTE: The numbers above indicate the mounting 
screw torque sequence: pre‐torque to 1.1 N‐M ± 0.2 N‐
Mounting
M (10 ± 2 inch‐lbs), then torque to 2.5 N‐M ± 0.2 N‐
Screws
M (22 ± 2 inch‐lbs).
FIGURE 20.1 Replacing 4-Way Valve and/or Manifold Assembly
67
Part No 1154245