SECTION 6—PREVENTIVE MAINTENANCE
PREVENTIVE MAINTENANCE RECORD INVACARE PERFECTO
Model No. IRC _______________________________
Serial No.____________________________________
On Each Inspection
Record Date Of Service
Record Elapsed Hours on
Hour Meter
Clean Cabinet Filters
Check Prescribed L/min.
Flow Rate
Annually, During Preventive Maintenance Schedule, Or Between Patients
Clean/Replace Cabinet
Filters
Check outlet HEPA Filter*
Check Compressor Inlet
Filter*
Check Oxygen
Concentration (SensO
2
models)
Check Power Loss Alarm
As Required or as otherwise noted
Clean Heat Exchanger
Replace Exhaust Muffler
Rebuild Top End
Compressor
Every 180 Days
Check oxygen
Concentration
(Units without SensO2)
*NOTE: Refer to Preventive Maintenance section of Service Manual.
™
Perfecto
2
CONCENTRATOR
30
OXYGEN
2
Part No 1154245