Invacare IRC5PO2AW Service Manual page 37

Oxygen concentrator homefill compatible
Hide thumbs Also See for IRC5PO2AW:
Table of Contents

Advertisement

O-ring, Intake Valve Keeper and Valve Keeper Strip
NOTE: For this procedure, refer to FIGURE 7.5.
1. Remove the O‐ring, intake valve 
flapper and valve keeper strip from the 
bottom of the valve plate. Discard all.
2. Clean the bottom of the valve plate 
with a clean, soft cloth.
Exhaust Valve Flapper, Valve Restraint and Valve Keeper Strip
NOTE: For this procedure, refer to FIGURE 7.6.
1. Remove the exhaust valve flapper, 
valve restraint and valve keeper strip 
from the top of the valve plate. Discard 
all.
2. Clean the top of the plate with a clean, 
soft cloth.
Gasket
NOTE: For this procedure, refer to                                
FIGURE 7.6.
1. Remove the old gasket from the top of 
the valve plate and discard.
2. Clean the valve plate with a clean, soft 
cloth.
Part No 1154245
SECTION 7—COMPRESSOR
FIGURE 7.5 Removing Compressor and
Compressor Heads - Rebuilding the
Compressor - O-ring, Intake Valve Keeper
and Valve Keeper Strip
Gasket
FIGURE 7.6 Removing Compressor and
Compressor Heads - Rebuilding the
Compressor - Exhaust Valve Flapper, Valve
Restraint and Valve Keeper Strip
37
Valve Flapper
Screw
Valve
Keeper Strip
O-Ring
Intake Valve
Flapper
Valve Flapper
Screw
Valve
Keeper Strip
Exhaust Valve
Flapper
Valve
Restraint
Perfecto
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Irc5pawPerfecto2 irc5po2awPerfecto2 irc5paw

Table of Contents