Warning And Safety Instructions; Connecting The H Solution Container - Miele G 7836 CD Operating Instructions Manual

Washer-disinfector
Hide thumbs Also See for G 7836 CD:
Table of Contents

Advertisement

Warning and Safety
instructions
,
These warning and safety
instructions are in addition to those
given at the beginning of this
booklet.
H
O
solution must only be used in
2
2
the special containers and
adapters supplied by the Ecolab and
Dr. Weigert companies.
Important: observe the
manufacturer's safety instructions
as given on their safety data sheet.
Be especially careful when
handling H
O
2
irritant chemical.
Observe all current safety instructions
pertaining to its use.
Wear protective gloves and goggles.
Empty containers must be
disposed of in accordance with the
manufacturer's instructions.
H
O
solution must not be mixed
2
2
with other chemicals. This could
cause a serious chemical reaction, e. g.
the release of harmful vapours or
gases.
solution. It is an
2
Special features - OXIVARIO
Connecting the H
container
The connection hose for the H
solution container is identified by a
black label. It is supplied without an
adapter because the system for
drawing the solution out of the
container varies depending on supplier.
^ Connect the supplier's adapter to the
connection hose (black).
^ Then connect this to the H
solution container.
^ Then start the DOS2-FILL
programme.
Unlike with other chemical agents, the
H
O
solution container must be fully
2
2
emptied before it is exchanged for a
new one.
Do not replace the H
container with a new one until the FILL
DOS 2 CONTAINER message is
displayed. After connecting the new
container you must run the DOS2-FILL
programme.
If the message CHECK DISPENSING
SYSTEM 2 appears in the display you
must check the container and all
dispensing tubes and hoses. The
programme will stop automatically.
O
solution
2
2
O
2
2
O
2
2
O
solution
2
2