Cancelling a programme

A programme can only be cancelled
in operating level B or D.
When testing to EN ISO 15883 or
HTM 2030 standards, the
programme must not be interrupted
or cancelled.
In operating levels B or D
^ Press the 7 button. The programme
is interrupted.
^ If the water temperature is below
40 °C, üCANCEL< or üCONTINUE<,
will appear in the display, and if it is
above 40 °C üCANCEL< will appear
in the display.
,
With established
decontamination programmes the
contaminated water must be
decontaminated before it is
discharged into the sewerage
system. Only then can the door be
opened.
If a great deal of hot water is present
in the wash cabinet when the
programme is interrupted and the
door is then closed quickly, hot wa-
ter can escape.
Danger of burning or scalding.
38
^ Use cursor 1 to select üCANCEL<;
the ü < cursors will start flashing.
^ Press the Start button 6. The
programme is cancelled and the wa-
ter pumped away. DRAIN WATER
appears in the display.
^ After the water has been pumped
away the required programme can
be selected and started again.