Page 2
NDSsi. NDS Surgical Imaging, NDSsi, the NDSsi logo, Dome, the Dome logo, Dome CXtra, and DICal are either registered or unregistered trademarks of NDS Surgical Imaging.
Product Information The design of the GX2MP display takes into account every known measure to ensure your personal safety. Improper use of the display can result in electric shock, fire, or damage to components. Read all instructions before setting up the display, and observe the safety precautions listed.
Symbol explanations CAUTION. Read the accompanying text carefully, for proper operation and maintenance of the display system. DANGEROUS VOLTAGE. Important precautions about electric shock. Read the accompanying text carefully, to prevent damage to components of the display system and for your safety. DIRECT CURRENT.
• Keep the display dry in all instances. The display lacks protection against liquids or spills. • In locations where 240V outlets are used, connect the display to a center-tapped, 240V, single-phase supply only (for Canada and the United States). vi | GX2MP Display...
Unpacking and handling tips The GX2MP display requires proper care to maintain product operation and adherence to specification. Unpack the display and components carefully, then set up and handle the unit properly to avoid damage to the LCD panel. • Use both hands to grasp the display case when lifting it from the shipping carton, but avoid touching the screen.
• Do not use aerosol directly on the display screen. Overspraying may cause electric shock. WARNING DISCONNECT SUPPLY BEFORE SERVICING AVERTISSEMENT COUPER L’ALIMENTATION AVANT L’ENTRETIEN ET LE DEPANNAGE viii | GX2MP Display...
Shipping/storing the display Keep the display in its shipping container until installation. Return the display to its original container whenever you need to store the unit, move it to another location, or return it for repair. The packaging supplied by the manufacturer protects the display while it is in transit.
Page 10
GX2MP display mounted on desk stand Power cord DVI cable Display controller Display driver and Quick reference calibration software Also included: D-sub cable NOTE: Some boards requires a DisplayPort-to-DVI adapter for each connector port.
16.7 million colors. Bundled with the display system, the Dome CXtra software enhances the functionality of the GX2MP display with a range of value-added services, such as DICOM calibration, error reporting, and backlight saver.
LED to glow amber. See “Troubleshooting” on page 18 for more information. Legend 1 Onscreen menu controls. Adjust display settings. 2 Power button. Powers display ON or OFF. 3 Power indicator. Reports display status. 2 | GX2MP Display...
Page 13
2 DVI, digital video input port. Requires a DVI connector. 3 VGA, analog video input port. Requires a D-sub connector. NOTE: Dome supports the GX2MP display only in digital mode. Use of DICOM calibration with the Dome CXtra software can only be supported on the DVI-D interface.
Orientation mode To adjust the viewing angle, use To rotate the display, use both both hands to swivel the display hands to turn the head clockwise head from side to side. from landscape to portrait mode. 4 | GX2MP Display...
Desk Stand Features Unfasten the stand lock to adjust display height. Remove the stand cover to thread the power cord and video cable connections to the display. To activate the USB hub function, the display must be connected to a USB-compliant computer or another hub with a USB cable.
Also disable any onboard graphics capability on the motherboard. NOTE: Dome supports the GX2MP display only in digital mode. Use of DICOM calibration with the Dome CXtra software can only be supported on the DVI-D interface. Do not use analog mode; that is, the D-sub interface.
Dome CXtra software, use the instructions shipped with that graphics card. • Video mode. Although the GX2MP display can be used with either a digital or analog video cable, the installation instructions in this manual relate to the use of the display with the DVI cable. Use of...
Page 18
Make sure the cable and cord run through the notches. To reattach the stand cover, align the hooks with the slots on the stand. Press the cover into place. A click sound signals a secure connection. 8 | GX2MP Display...
Install the Display Driver Before you install the driver, remove any previously installed display driver for the display controller from your system. After installation, use the Display Properties dialog to makes changes to the video settings or to set up a dual-display configuration. To install the driver 1 Start the system.
Page 20
1 Right-click the desktop and select Properties > Settings. 2 Click the monitor icon 2. Then select Extend my Windows desktop onto this monitor. 3 Set the resolution and color quality for the second display. 4 Click Apply or OK. 10 | GX2MP Display...
Using Display Controls Use the onscreen menu to make adjustments to display settings such as brightness, contrast, position, and language. (See the table on page 14 for the functions available by video mode.) You can access the menu whenever both the computer and display are on. If the computer is off or in power-saving mode, the menu is inaccessible.
Page 22
Self Image Setting function is executed automatically.) POWER INDICATOR Lights up blue when the display operates normally (ON mode). In Sleep (energy- saving) mode, the color changes to amber. In OFF mode, no color appears. 12 | GX2MP Display...
Screen Adjustment IMPORTANT: Allow the display to stabilize for at least 30 minutes before making image adjustments. This illustration shows the logic of the onscreen menu and relates to the following procedure. To make screen adjustments 1 Press the MENU button. The main menu appears. 2 Use the MINUS or PLUS button to access a control.
Page 24
Language Customizes the screen status for the operating environment. Horizontal Position Vertical White Balance Power Indicator Factory Reset f-ENGINE Movie/Test Selects or customizes desired image settings User Normal A - Analog; D - Digital; X- Adjustable 14 | GX2MP Display...
Main Menu Main Menu Submenu Description Brightness Adjust the brightness of the screen. Contrast Adjust the contrast of the screen. Gamma Set gamma value: -50/0/50. On the display, high gamma values display whitish images and low gamma values display high contrast images. Preset sRGB: Set the screen color to fit the SRGB standard color...
Set indicator to ON or OFF. If you set OFF, it will go off. If you set ON at any time, the power indicator automatically turns on. Factory Reset Restore all factory default settings except LANGUAGE. Press PLUS to reset immediately. 16 | GX2MP Display...
This table shows the additional commands available via the f-ENGINE menu. NOTE: Changing screen settings, such as brightness, via the f-ENGINE menu voids DICOM calibration on the GX2MP display. Do not use this menu to change settings. Main Menu Submenu...
Check the specification of your video card. monitor, Plug & Play support Plug and Play Replace the card if it does not support the (VESA DDC) monitor function. Plug and Play function. found” message on the screen. 18 | GX2MP Display...
Page 29
Problem Possible Cause Solution Display image is Display position is Press AUTO/SET to automatically adjust incorrect. incorrect. your display image to the ideal setting. If the results are unsatisfactory, adjust the image position using the H position and V position icon in the on screen display. Check Control Panel >...
Page 30
445 mm x 345 mm x 72 mm (17.5 x 13.8 x. 2.8 in) Weight With stand Less than 10 kg (23 lb) Mounting options Standard 100 mm VESA mounting standard Desktop stand Height, pivot, swivel, and tilt adjust 20 | GX2MP Display...
Page 31
Environment Category Item Characteristic Item Specification Temperature operating 0° to 40° C storage -10° to 60° C Humidity operating 10% to 80% Relative Humidity (noncondensing) storage 5% to 90% Relative Humidity (noncondensing) Power Indicator Mode LED Color Blue Sleep Amber None Technical Information | 21...
Page 33
DVI Connector Port The video signal connector is a standard DVI connector. Signal (DVI-D) Signal (DVI-D) T.M.D.S.*Data2- T.M.D.S. Data3+ T.M.D.S. Data2+ +5V Power T.M.D.S. Data2/4 Shield Ground** T.M.D.S. Data4- Hot Plug Detect T.M.D.S. Data4+ T.M.D.S. Data0- DDC Clock T.M.D.S. Data0+ DDC Data T.M.D.S.
Screen Saver) automatically turns the backlight off during the period of inactivity. Backlight Saver and Screen Saver both extend the life of the backlight and reduce burn-in of images. Do not use the Backlight Saver and Screen Saver concurrently. 24 | GX2MP Display...
Index adjusting screen managing power usage 11, 12, 14, 15, 17 menu, onscreen viii cleaning display components, display package contents connecting display power cord, connecting connector port, DVI power management contents, package precautions controls, display product about viii cleaning desk stand disposal display information...
Regulatory Compliance U.S. FCC Compliance Statement This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Safety Certification Canada and the United States European Union Germany TÜV Rheinland GS Mark China Compulsory Certification Japan Voluntary Control Council for Interference Korea Communications Commission...
Standard Warranty Return Procedure As a NDSsi Standard Warranty customer, you must follow the procedure below if you have a nonfunctioning GX2MP display. The Customer Service staff will attempt to correct any minor issues that may be causing the problem. Once NDSsi has determined that you have a nonfunctioning product, we will arrange for return and repair of the nonfunctioning product.
Page 39
e) Causes external to the product, such as electric power fluctuations or failure. f ) Use of supplies or parts not meeting the NDSsi specifications. g) Normal wear and tear. h) Any other cause, which does not relate to a product defect. 4 Warranty excludes removal, installation, and setup service charges.
Page 40
NDS Surgical Imaging, LLC 5750 Hellyer Avenue San Jose CA 95138-1000 USA www.ndssi.com...