Selecting A Range - Agilent Technologies U3606A User's And Service Manual

Multimeter - dc power supply
Table of Contents

Advertisement

2
Digital Multimeter Operation

Selecting a Range

Selecting a Range

You can allow the U3606A to automatically select the range using
autoranging, or you can select a fixed range using manual ranging.
Autoranging is convenient because the U3606A automatically selects the
appropriate range for sensing and displaying each measurement. However,
manual ranging results in better performance, since the U3606A does not
have to determine which range to use for each measurement.
Key
• Autoranging is selected at default factory power- on and after a remote
• Manual ranging — If the input signal is greater than can be measured
• For frequency, pulse width, and duty cycle measurements, ranging
• The range is fixed for diode (1 V
• The U3606A remembers the selected ranging method (auto or manual)
• Autorange thresholds — The U3606A shifts ranges as follows:
54
Description
Press
Press
Press Shift >
reset.
on the selected range, the multimeter provides these overload
indications: "OL" from the front panel or "±9.9E+37" from the remote
interface.
applies to the input voltage or input current of the signal, not its
frequency.
and the selected manual range for each measurement function.
Down range at < 10% of current range
Up range at > 120% of current range
to select a higher range and disable autoranging.
to select a lower range and disable autoranging.
Auto
to enable autoranging and disable manual ranging.
range) tests.
dc
U3606A User's and Service Guide