Attach the epi-fluorescence cube turret and epi-fluorescence attachment (required for
6
epi-fluorescence microscopy).
Tool: Hex driver (2 mm across flats)
Hex wrench (3 mm across flats)
First, connect the epi-fluorescence cube turret and
epi-fluorescence attachment and then attach them on the
microscope arm.
(1)
Insert the positioning pin of the epi-fluorescence
attachment while aligning it with the positioning hole
on the epi-fluorescence cube turret side, and then
tighten the connection fixing screws (x2) on the
epi-fluorescence cube turret with a hex driver.
(2)
When using a motorized or intelligent
epi-fluorescence cube turret, remove the contact
cover on the contact arm beforehand.
(Contact cover is not removed for manual
epi-fluorescence cube turret.)
(3)
Loosen the fixing screw on the front of the arm using
a hex driver so that the tip of the screw does not
protrude into the connecting section.
(4)
Align the round dovetail and convex contact on the
bottom of the epi-fluorescence cube turret with the
round dovetail and concave contact of the arm and
slide the entire mounted attachment to the rear.
(There is no concave contact in manual
epi-fluorescence cube turret.)
(5)
Tighten the fixing screw loosened in step (3).
(6)
Tighten the four fixing screws on top of the
epi-fluorescence attachment.
(7)
Attach the HG adapter to the bayonet mount at the
rear of the epi-fluorescence attachment and connect
the HG precentered fiber illuminator. (See the
instruction manual provided with the HG
precentered fiber illuminator.)
Chapter 1
Assembly
Connection fixing screw
Connecting epi-fluorescence cube turret
2
Fixing screw
Attaching and fixing epi-fluorescence cube turret
F. ST
OP
A. ST
7
Epi-fluorescence
Positioning hole
cube turret
F. ST
OP
A. ST
OP
and epi-fluorescence attachment
A. ST OP
F. ST OP
1
2
3
4
5
6
and epi-fluorescence attachment set
Bayonet mount
HG adapter
OP
Attaching the HG adapter
Epi-fluorescence
attachment
Epi-fluorescence
attachment
fixing screw
HG fiber