Injections are labeled as RUN 1, 2, etc. Subsequent injections are completed once START is
redisplayed. A total of 6 RUNS can be displayed. Further RUNS require deletion of one of
the 6.
11. Press the VIEW RUN 1 (2-6) softkey to view injection. To delete the run, press DEL RUN 1
(2-6) softkey and press the YES softkey.
12. Press the AVG CO softkey after all runs to be computed in the average are present. The
preliminary hemodynamic table is displayed.
NOTE:
Injectate temperature (Ti), and blood temperature (Tb) are
displayed in the CO trace. There must be at least a 12° C difference
between Ti and Tb. When using iced solution or a CO-SET
remaining fluid in the line— the Ti measured by the in-line probe will
be more than 12° C cooler than Tb at the time of the injection
regardless of the value of Ti displayed prior to the injection.
13. Vital sign parameter data that may require editing is displayed in right hand column. Review
these values.
To edit these values:
Press the EDIT VS softkey
Press the ADJ ↑↓ softkeys to adjust highlighted value
Press the NEXT ↑↓ → softkeys to move cursor to next vital sign for editing
14. Press the RUN CALCS softkey to display altered preliminary hemodynamic table.
15. Press the STORE CALCS softkey to display completed hemodynamic table.
16. Press the INDEX ON/OFF softkey to toggle between indexed and nonindexed values.
17. Hemodynamic table can be recorded with a bedside recorder. While hemodynamic table is
displayed on monitor screen, press the RECORD key and within 2 seconds press the softkey
next to CO label.
CO allows 22 sets of hemodynamic data to be stored.
18. To view hemodynamic table at a later time, press the CO softkey, press the MORE softkey,
and press HEMOCALCS. Use softkey arrows to scroll table information.

Selecting a Thermodilution Catheter

The following Baxter-Edwards models of thermodilution catheters or their equivalent are
recommended. "H" in part number refers to Heparin coating, and can be ordered without Heparin
by excluding the "H" from part number.
Part Number
131HF7
831HF75
93A-931H-7.5F
93A-439H-7.5F
Standard, non-REF capable
Standard, non-REF capable with venous infusion port
Non-REF capable, infusion and pacing
REF capable, volumetric
3-18
Application
®
— due to