Storage Of Invasive Pressure Values To Vsc; Th Trace Option - Escort Prism 20401 Reference Manual

Patient monitor
Table of Contents

Advertisement

4. Press the CURSOR ↑↓ softkeys to align cursor bar with end expiration. Note value displayed
on cursor bar and at bottom of screen changes as cursor bar is moved.
5. Press the SAVE PAWP softkey to store PAWP value for hemodynamic calculations. Press
the FREEZE key to release (if FREEZE key was used to hold display).
6. For a PA catheter that is not to be wedged, the monitor can store pulmonary artery diastolic
value in place of PAWP. Press the PAWP=CURSOR/PAD softkey to select PAD.
7. If a separate CVP line is not present, CVP can be temporarily measured by "bridging PA
line" to determine CVP value. Press the PA softkey, press MORE, then press SAVE
PAM;CVP softkey.
NOTE: Storage of PAWP and CVP values requires initiation of
Cardiac Output injections within 30 minutes for these values to be
used for hemodynamic calculations.
Storgae of Invasive Pressure Values to VSC
STORAGE OF INVASIVE PRESSURE VALUES TO VSC: Invasive pressure values are
dynamic values that store into VSC any time a static value such as NIBP or SureTemp is stored.
A single set can be stored into VSC through STORE VS function. These dynamic values can be
stored at a timed interval as follows:
1. Press the softkey next to the VSC label to open the VSC SETUP menu.
VSC
STORE
TABLE
→ →
SETUP
VS
2. Press the VSC SETUP softkey to display the following menu.
AUTO
STORE
OFF
ALM
1MIN
SETUP
3. Press the AUTO softkey to highlight the desired storage interval.
4. Press the PAGE HOME key.
until user returns to this menu and makes another selection.
th
5
Trace Option
If the Prism or Prism SE is configured with the 5
can be split to display 2 Invasive Blood Pressures. The selections for displaying a 5
TRACE 4 are BP1 & 2, BP1 & BP3, and BP2 & BP3. The waveform label for TRACE 4 will
display as IBP.
1. To access the individual BP menu, press the IBP softkey, a submenu will appear. Press the
desired BP to access the associated BP menu.
2. To record, press the appropriate BP softkey during the tab recording to designate that
parameter for recording.
TABLE
TABLE
TEXT
↑ ↑
↓ ↓
SIZE
STORE
STORE
STORE
NIBP
TMP1
CO
ON
ON
ON
Timed storage continues until monitor is turned off or
th
Trace Option, TRACE 4 (Waveform Zone 4)
3-16
MORE
MORE
th
waveform in