Spo2 Messages; Invasive Blood Pressure (Bp1/Bp2/Bp3) Messages - Escort Prism 20401 Reference Manual

Patient monitor
Table of Contents

Advertisement

SPO2 Messages

SPO2 MESSAGE
SPO2 SRCH
(Nellcor only)
NO C-LOCK
(Nellcor only)
NO SENSOR
SPO2 CAL

Invasive Blood Pressure (BP1/BP2/BP3) Messages

BP1/BP2/BP3 MESSAGE
NO XDUCER
ZERO BP1
ZERO BP2
ZERO BP3
NO ZERO: PULSE DETECT
ZERO OUT OF RANGE
The monitor cannot locate the patient's pulse. The patient's
perfusion may be too poor to detect an acceptable pulse. Confirm
proper application of the sensor; make sure the ECG parameter is
available for C-LOCK synchronization; try another sensor site; or
try the OXISENSOR II R-15 sensor.
Indicates loss of synchronization between the ECG and SPO2
pulse waveform. Take steps to eliminate ECG artifact, or if the R-
wave amplitude is less than 4 mm, increase the ECG waveform
size by using the ECG SIZE softkey.
The SPO2 parameter is ON, but the sensor and/or sensor cable
is/are not connected to the monitor.
The monitor is performing automatic self-calibration. The
calibration range is 50 to 100%.
The pressure transducer is not connected to the monitor.
Message flashes until the pressure channel is zeroed as a
reminder that pressures must be zeroed prior to use.
Displayed over the BP SETUP page to indicate that either a pulse
or electrical noise was detected, which precluded zeroing the
transducer. Confirm that the transducer is vented to air before
pressing the ZERO key.
Displayed over the BP SETUP page to indicate that a
transducer's offset exceeds the zero range (±100 mmHg) of the
monitor. Confirm that the transducer is vented to the air and
attempt to zero again.
7-3
MEANING
MEANING