Cardiac Output (CO) Messages

CO MESSAGE
NO CATH
NO PROBE
∧ ∧ RANGE ∧ ∧
∨ ∨ RANGE ∨ ∨
<< RANGE >>
BASELINE
IRG CURVE
PEAKS
INJ LATE
DRIFT
SHORT CUR
LONG CURV
WARM H2O
NO ECG
<<EF HR>>
IRG ECG
<<CO EF>>
FAST DROP
EF EARLY
EF LATE
<<BOLUS>>
RUNS FULL
DELETED
No catheter connected to the monitor.
No temperature probe (bath or injectate) connected to the
monitor.
Signal over range.
Signal under range.
Cardiac output out of range.
Noisy baseline detected.
Irregular curve detected.
Multiple curve peaks detected.
Delayed injection detected.
Excessive thermal drift detected.
Short thermodilution curve detected.
Long thermodilution curve detected.
Warm H2O for injectate or bath detected.
No ECG signal detected by the monitor.
REF HR out of range.
Irregular rhythm.
REF out of range.
Fast drop curve.
Premature beat.
Delayed beat.
REF post bolus out of range.
Maximum number of runs performed.
Measurement run deleted.
7-6