Invacare Platinum 5 Service Manual page 78

Platinum series 5 liter oxygen concentrator
Hide thumbs Also See for Platinum 5:
Table of Contents

Advertisement

SECTION 20—4-WAY VALVE
13. Cut the tie-wraps that secure the Y-tube muffler assembly to the manifold assembly.
14. Remove the Y-tube muffler assembly from the manifold assembly. Use a screwdriver
to disengage the tubing connection at the barbed fitting while pulling the tubing away
from the manifold assembly.
Sieve Bed Hoses
NOTE: Manifold assembly
shown is for 10 liter models.
4-Way Valve
Valve Manifold Input
Hose
NOTE: New Heat Exchanger
does not have fins.
Heat Exchanger
Compressor Output
Hose (BLUE)
FIGURE 20.3 Replacing Manifold Assembly (Units Manufactured after 7/24/03)
Platinum™Series
Tie-Wrap
Muffler
without
Throttling
Insert
Adjustable
Clamp
78
Spade Connectors
Tie-Wraps
Valve Pilot
Manifold Assembly
Bottom Center Port
Adjustment Clamp
Y-tube Muffler Assembly
DETAIL "A" -
MANIFOLD ASSEMBLY
Mounting
(Two [2]
per side)
Sound Box
Top
Part No 1110538
Screw
Bushing