Invacare Platinum 5 Service Manual page 63

Platinum series 5 liter oxygen concentrator
Hide thumbs Also See for Platinum 5:
Table of Contents

Advertisement

11. Remove spade connectors from circuit breaker, hour meter and On/Off (
(Jumper wire harness from On/Off switch to circuit breaker does not have to be
removed).
12. Remove control panel and reverse STEPS 3-10 to re-install control panel.
13. Run unit and check for leaks at pressure sensor tubing on product tank cap. Refer to
Leak Test on page 89
14. Reinstall cabinet. Refer to Removing Cabinet on page 25.
Hour Meter
(STEP 10)
1/8" I.D. Pressure
Sensor Tubing (STEP 5)
Oxygen Sensor Tube
(SensO
Models Only)
2
Transformer
Harness
Connector
(STEP 10)
9 or 10-Pin
Connector
(STEP 8)
8-Pin
Connector
(STEP 9)
Part No 1110538
Circuit Breaker
(STEP 10)
Lower Mounting Screws
FIGURE 13.1 Removing Control Panel
SECTION 13—CONTROL PANEL
63
) switch.
Upper Mounting
Screws
ON/OFF
(
)
Switch
(STEP 11)
Patient Outlet
Barbed fitting
Top and Bottom
Barbed fittings
Control Panel
Platinum™Series

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Platinum 10Platinum xl

Table of Contents