SECTION 4—TROUBLESHOOTING
SYMPTOM
LOW PRESSURE:
Internal Status Indicators:
RED: One Flash
GREEN: One Flash
Or
RED: One Flash
GREEN: Three Flashes
Control Panel Indicators:
RED: On
YELLOW: Off
GREEN: Off
Unit plugged in, power
switch on, continuous
audible alarm.
Compressor shut down
(Failure to cycle due to
low pressure).
Defective Parts:
Internal Status Indicators:
RED: One Flash GREEN:
Five Flashes
Or
RED: Two Flashes
GREEN: Three Flashes
CONTROL PANEL
INDICATORS:
RED: On
YELLOW: Off
GREEN: Off
Unit plugged in, power
switch on, continuous
audible alarm.
Compressor shut down.
Platinum™Series
PROBABLE CAUSE
Compressor inlet filter dirty or plugged.
Compressor:
a. Leaks at fittings or tubing
b. Leaking or defective relief valve.
c. Insufficient voltage at outlet.
d. Worn cup seals or gaskets.
Heat exchanger:
a. Leak at tubing or body chamber.
b. Inspect tubing and heat exchanger.
Regulator cracked or leaking.
P.E. valve leaking.
Defective main valve coil.
Connector loose.
Defective P.E. valve coil. (Resistance 80
ohm's ± 10.)
Defective P.E. valve.
Connector loose.
Defective P.C. board.
Replace compressor inlet filter. Refer to
Replacing the Compressor Inlet Filter on
page 28.
a. Replace or repair.
b. Repair leak or replace.
c. DO NOT use extension cords. Use
another outlet.
d. Rebuild top end of compressor. Refer
to Rebuilding the Thomas Model
2650 and 2660 Compressor -
Platinum 5 and 10 Models on page 37
or Platinum XL Model - Rebuilding
the Gast Compressor on page 45.
a. Replace or retighten.
b. Replace or retighten tubing. Replace
heat exchanger. Refer to Replacing
Heat Exchanger on page 61.
Replace regulator. Refer to Replacing
Regulator on page 58.
Replace P.E. Valve. Check voltage at P.E.
valve connector on 24 volt D.C. scale. The
P.E. Valve activates or energizes
approximately one second prior to the
activation of the 4 way valve with
approximately 24 volts. If voltage is in
excess of 24 volts consistently, replace the
P.C. board. Refer to Replacing P.C. Board
on page 66. If the P.C. board voltage acts
normally, replace the P.E. valve. Refer to
Replacing P.E. Valve on page 53.
NOTE: Check for leaks starting at the
compressor output through all the
pneumatic connections. Major leaks will
cause system pressures to remain below
adequate shift (exhaust) pressures and will
cause compressor shutdown.
a. Replace main valve.
b. Replace main valve coil.
Reattach connector.
Replace P.E. coil if shorted or open on
resistance check. Refer to Replacing P.E.
Valve on page 53.
Replace P.E. valve. Refer to Replacing P.E.
Valve on page 53.
Reattach connector.
Replace P.C. board. Refer to Replacing P.C.
Board on page 66.
20
SOLUTION
Part No 1110538