Invacare Platinum 5 Service Manual page 32

Platinum series 5 liter oxygen concentrator
Hide thumbs Also See for Platinum 5:
Table of Contents

Advertisement

SECTION 6—PREVENTIVE MAINTENANCE
PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN
Model No. IRC _______________________________
Serial No.____________________________________
On Each Inspection
Record Date Of Service
Record Elapsed Hours on
Hour Meter
Clean Cabinet Filters
Check Prescribed L/min.
Flow Rate
Every 180 Days FOR IRC5LX (EVERY 90 DAYS FOR IRC10LX)
Check oxygen
Concentration
(Units without SensO2)
Annually, During Preventive Maintenance Schedule, Or Between Patients
Clean/Replace Cabinet
Filters
Check outlet HEPA Filter*
Check Compressor Inlet
Filter*
Check Oxygen
Concentration (SensO
2
models)
Check Power Loss Alarm
As Required
Clean Heat Exchanger
Replace Exhaust Muffler
Rebuild Top End
Compressor
*NOTE: Refer to Preventive Maintenance section of Service Manual.
Platinum™Series
CONCENTRATOR
32
Part No 1110538

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Platinum 10Platinum xl