Invacare Platinum 5 Service Manual page 47

Platinum series 5 liter oxygen concentrator
Hide thumbs Also See for Platinum 5:
Table of Contents

Advertisement

2. Turn the valve plate assembly over.
When removing the old O-ring gaskets ensure not to damage the grooves they sit in.
3. Using a small screwdriver, remove the old O-ring gaskets from the valve plate assembly.
4. Using a ¼-inch socket or nut driver, remove the screw, the valve retainer and intake
valves from the valve plate assembly.
5. Turn the valve plate assembly over.
6. Using a ¼-inch socket or nut driver, remove the screw, old exhaust valves and limiters
of the valve plate chambers.
Part No 1110538
CAUTION
FIGURE 7.12 Removing the Old O-Ring Gaskets
¼-inch Socket or
Nut Driver
Intake Valves
FIGURE 7.13 Removing the Intake Valves
O-Ring Gaskets
Valve Plate
Assembly
Screw
47
SECTION 7—COMPRESSOR
Valve Plate
Assembly
Valve
Retainer
Platinum™Series