Invacare Platinum 5 Service Manual page 30

Platinum series 5 liter oxygen concentrator
Hide thumbs Also See for Platinum 5:
Table of Contents

Advertisement

SECTION 6—PREVENTIVE MAINTENANCE
3. Remove the two mounting screws from rear of sound box.
4. Disconnect and remove the manifold exhaust tubing from the barbed fittings on top
of the exhaust canister.
5. Unscrew the exhaust muffler located on the exhaust canister by turning
counterclockwise.
6. Inspect exhaust canister to ensure the barbed fittings are clear and do one of the
following:
• Replace exhaust canister and muffler if plugged or restricted.
• Install only new exhaust muffler if exhaust canister is in good condition.
7. Install new/re-install existing exhaust canister by reversing STEPS 3-5.
8. Reinstall cabinet. Refer to Removing Cabinet on page 25.
Exhaust Canister
Exhaust Muffler
FIGURE 6.5 Replacing the Exhaust Canister/Exhaust Muffler (Platinum 5 Models Manufactured
Platinum™Series
Manifold Exhaust Tubing
before 7/24/03)
30
Barbed Fittings
Mounting Screw(s)
Rear of Sound Box
Part No 1110538

Hide quick links: