ResMed Stellar 100 User Manual page 32

Invasive and noninvasive ventilators
Hide thumbs Also See for Stellar 100:
Table of Contents

Advertisement

Sound pressure level
Alarm Volume Range
Dimensions (L x W x H)
Weight
Air outlet
Pressure measurement
Flow measurement
Power supply
External DC Power
Supply
Internal Battery
Housing construction
Environmental
conditions
Electromagnetic
compatibility
Air filter
Air tubing
SlimLine air tubing
IEC 60601-1
classifications
29 dBA as measured according to ISO 17510 – 1; 2002.
32 dBA with uncertainty of 3 dBA as measured according
to ISO 17510 – 1:2007 .
> 45 dBA - <85 dBA at 3' 3" (1 m); 3 steps: low medium,
high
9.1" x 6.7" x 4.7" (230 mm x 170 mm x 120 mm)
4.6 lb (2.1 kg)
0.9" (22 mm) taper, compatible with ISO 5356-1:2004
Anaesthetic & Respiratory Equipment - Conical Connectors
Internally mounted pressure transducer
Internally mounted flow transducer
AC 100–240V, 50–60Hz, 2.2 A, max. 65 W
24 V, 3 A
Lithium-Ion battery, 14.4 V, 1.6 Ah, 23 Wh
Operating hours: 2 h with a new battery under normal
conditions (see below).
Patient type: home chronic; pressure: IPAP/EPAP 15/5
cm H
O; mask type: Ultra Mirage; air tubing: 6' 6" (2 m);
2
leak: 0; respiratory rate: 20 bpm; battery capacity: 100%
Patient type: hospital acute; pressure: IPAP/EPAP 20/5
cm H
O; mask type: Ultra Mirage; air tubing: 6' 6" (2 m);
2
leak: 0; respiratory rate: 45 bpm; battery capacity: 100%
Flame retardant engineering thermoplastic
• Operating temperature: +32°F (0°C) to + 95°F (35°C)
• Operating humidity: 10%–95% non-condensing
• Storage and transport temperature: -4°F (-20°C) to
+140°F (60°C); +122°F (+50°C*)
• Storage and transport humidity: 10%–95% non-
condensing
• Air pressure: 1,100hPa to 680hPa; Altitude: 11482' 9"
(3,500 m)
*NONIN XPOD
Product complies with all applicable electromagnetic
compatibility requirements (EMC) according to
IEC60601-1-2, for residential, commercial, and light
industry environments. For further details see "Guidance
and manufacturer's declaration – electromagnetic
emissions and immunity" on page 28.
Electro static fibre mesh with TPE frame structure. Bacterial
filtration efficiency of 99.540% on area weight 100g/m².
Flexible plastic, 6' 6" (2 m) or 9' 10" (3 m) length
(0.9"/22 mm diameter)
Flexible plastic, 6' (1.83 m) length (0.6"/15 mm diameter)
• Class II (Clause 3.14—double insulation). This
adherence means the need for an protective earthing
(ie, an earthed plug) is not necessary.
• Type BF
• Continuous operation
Technical specifications
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Stellar 150

Table of Contents