ResMed Stellar 100 User Manual page 11

Invasive and noninvasive ventilators
Hide thumbs Also See for Stellar 100:
Table of Contents

Advertisement

1
Connect the power cord.
2
Plug the free end of the power cord into a power outlet.
3
Connect the antibacterial filter firmly onto the air outlet of the device.
4
Connect the external humidifier to the other side of the antibacterial filter.
*If not using an external humidifier, the HMEF can be connected to the leak port (at step 9).
5
Connect the air tubing to the external humidifier.
6
Connect the leak port to the air tubing.
7
Select the mask type Trach (select Setup menu, then Clinical Settings, then Advanced
Settings).
8
Perform Learn Circuit (see "Setup menu: Options" on page 16).
9
If an external humidifier is not being used, you can connect the HMEF to the leak port if
necessary.
10
Connect the catheter mount.
** The leak port or the HMEF can be connected to standardized tracheostomy interfaces
including connector pieces like catheter mounts.
The catheter mount and the external humidifier are not part of the ResMed component.
6
Leak Circuit
configuration
1
AC power cord
AC locking clip
2
Antibacterial filter
5
3
4
External humidifier*
Catheter mount**
ResMed leak port
HMEF*
6
Air tubing
10