GasAlertMicroClip XT
Technical Reference Guide
Problem
Detector Operation
Detector does not display expected
gas readings after activation self-test.
Detector does not respond to
pushbutton.
Detector does not accurately measure
gas.
Detector does not enter into alarm.
54
Table 14. Troubleshooting
Possible Cause
Sensor not stabilized
Sensor(s) requires calibration
Target gas is present
Battery is depleted
Detector is performing operations that do
not require user input
Sensor(s) requires calibration
Detector is colder/hotter than ambient gas
Sensor filter is blocked
Alarm setpoint(s) defined incorrectly
Alarm setpoint(s) set to zero
Detector is in calibration mode
Solution
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
Calibrate the sensor(s). Refer to Calibration.
Detector is operating properly. Use caution in
suspect areas.
Charge battery. Refer to
Charging the
Pushbutton operation restores automatically
when the operation ends.
Calibrate the sensors. Refer to Calibration.
Allow the detector to attain ambient tempera-
ture before use.
Replace the sensor filter. Refer to
Sensor or Sensor
Filter.
Reset alarm setpoints. Refer to
Alarm Setpoints
and
Sensor
Reset alarm setpoints. Refer to
Alarm Setpoints
and
Sensor
Complete the calibration procedure.
Battery.
Replacing a
Factory Gas
Configuration.
Factory Gas
Configuration.