Selectable Weight Units and Piece Unit:
Kilogram
Gram
Pound
Pound/Ounce (lb, oz)
Ounce

Error Messages:

The EEPROM memory is not working correctly.
Zero is higher than the zero range when switching the scale on.
Zero is lower than the zero range when switching the scale on.
A/D value is unstable. When switching the scale on or pressing the
or the
A/D value is less than -31250.
A/D IC malfunction (cannot read A/D value). The load cell may not be connected to
the scale correctly.
A/D value is higher than +31250.
The weight of the object is over the maximum capacity + 9 divisions.
After calibration the displayed resolution is greater than the internal resolution.
The gravity value is outside the range 9.772 ~ 9.829.
Error setting.
SK-VET V1.0
(kg)
(g)
(lb)
(oz)
key, the A/D value is unstable for at least 20 seconds.
TARE
The weight value is greater than the zero range, when re-zeroing the
scale.
The weight value is below the zero range, when re-zeroing the scale.
.
1 g = 0.001 kg
1 g = 1 g
1 g = 0.002204623 lb
1 g = 0.03527396 oz
1 g = 0.03527396 oz