Power Description; Error Messages - Excell KWH3 User Manual

Weighing scale
Hide thumbs Also See for KWH3:
Table of Contents

Advertisement

EXCELL PRECISION CO., LTD.

1-4 Power Description

Power Selection
1. 6 V / 4 Ah Rechargeable battery
2. 100~240 V AC
Power Consumption
Indicator + L/C + no backlight
indicator+ L/C + front display
backlight
Low Power Alarm
Please note when the (
display, the batteries should be recharged.
The scale will turn off automatically after a few hours when the low battery warning symbol
shows up. The scale must be fully recharged, before operating again.
Please recharge at once when the symbol
accuracy.

1-5 Error Messages

 Weight exceeds 9d of maximum capacity. (d=division)
 Zero point positive deviation after power on exceeds +10% FS (i.e. zero point
deviation over 10% of full scale [max. capacity] above calibration zero point).
 Zero point negative deviation after power on exceeds -10% FS (i.e. zero point
deviation over 10% of full scale [max. capacity] below calibration zero point).
 Unstable zero return, unstable over 10 sec. Press
 Zero is too high when calibrating. (over internal value350,000)
 Zero is too low when calibrating. (under internal value 80,000)
-----
If there is no Tare or Pre-tare, the weight is less than -20d.
02029018
27 mA
37 mA
) symbol keeps flashing on the left down corner of the
shows in order to keep the weight
7
to leave E4.
ZSME300000363

Hide quick links: