6. Daily Inspection and Maintenance

6.3.2 Forehead Rest

CAUTION
Disinfect the forehead rest whenever changing patients.

6.3.3 Cover and Monitor

WARNING
!
Note: Do not clean the cover of the instrument with lens cleaner. The cover of the instrument
could be damaged.
Use the procedure below to clean the cover when it is dirty.
1
Turn OFF the power of the main unit.
Press the power switch to the O side.
2
Unplug the power cable.
3
Wipe the instrument with cleanser.
Soak a soft cloth with diluted neutral cleanser and wring it out.
4
Wipe off with a rinsed cloth.
Wipe with a cloth that has been dipped in water and wringed out.
To prevent the risk of infection, wipe the forehead rest with disinfecting
ethanol, glutaraldehyde solution, or other disinfectant whenever changing
patients.
For more details on the disinfection procedure, please consult a specialist.
Please also consult a specialist when using a disinfectant other than those
specified above and then adding another disinfectant to ethanol since this
could damage the forehead rest.
When the instrument is going to be cleaned, be sure to turn OFF the
power, and unplug the power cable from the AC outlet.
Never use alcohol, benzine, thinner or any other flammable cleaning
agents. Otherwise, fire or electric shock may result.
-30-