4. Operation
3
Unlock the stage unit.
Press the stage unit lock button to release the lock.
4
Adjust the height of the main unit.
Turn the operation lever, and align the main unit with the center position of the height adjustment
mark.
5
Darken the room.
Darken the room so that there is barely enough light (approx. 5 lux) by which to read a newspaper.
6
Input the study information into the computer.
Use the control software to input the study information.
7
Disinfect the face rest, and replace the chin rest paper.
CAUTION
!
CAUTION
!
To prevent the risk of infection, wipe the forehead rest with a disinfectant
such as ethanol or glutaraldehyde whenever the patient is changed.
For details on the disinfection procedure, consult a specialist.
To ensure cleanliness, replace the chin rest paper whenever the patient is
changed.
If the chin rest papers will not be used, disinfect the chin rest as well.
-10-
Center of height
adjustment mark