Fluke 1625-2 User Manual page 54

Earth/ground tester
Table of Contents

Advertisement

1625-2
Users Manual
Resistance Measurement (R
Measuring method:
Measuring voltage:
Short circuit current:
Measuring frequency:
Measuring Range
0.020 Ω...300 kΩ
Measuring time:
Maximum interference
voltage:
Maximum overload:
Resistance Measurement (R
Open circuit voltage:
Short circuit current:
Formation of measured value:
Measuring Range
0.020 Ω...3 kΩ
46
)
~
current and voltage measurement
20 V ac, square pulse
>250 mA ac
94, 105, 111, 128 Hz selected manually or automatically (AFC)
Display Range
Resolution
0.001 Ω ... 2.999 Ω
0.001 Ω
3.0 Ω ... 29.99 Ω
0.01 Ω
30 Ω ... 299.9 Ω
0.1 Ω
300 Ω ... 2999 Ω
1 Ω
3.0 kΩ ... 29.99 kΩ
10 Ω
30.0 kΩ ... 299.9 kΩ
100 Ω
typical 6 seconds
24 V, with higher voltages measurement will not be started
U
maximum = 250 V
rms
)
20 V dc
200 mA dc
with 4-pole measurement wires on  can be
extended without additional error
Resistances >1 Ω in wire  can cause additional error of
5 mΩ/Ω
Display Range
Resolution
0.001 Ω ... 2.999 Ω
0.001 Ω
3.0 Ω ... 29.99 Ω
0.01 Ω
30.0 Ω ... 299.9 Ω
0.1 Ω
300 Ω ... 2999 Ω
1 Ω
Accuracy
±(2 % of rdg +2 digit)
±(5% of rdg +5 digit)
Accuracy
±(2 % of rdg +2 digit)
±(5% of rdg +5 digit)
Operating Error
Operating Error

Advertisement

Table of Contents
loading

Table of Contents