ResMed VPAP III User Manual page 43

Bilevel pressure support ventilators
Hide thumbs Also See for VPAP III:
Table of Contents

Advertisement

Air Tubing: Flexible plastic, 2 m or 3 m length
IEC 60601-1 Classifications
Class II (double insulation)
Type CF
Table 7: Displayed values
Value
Pressure sensor at air outlet
Pressure
Flow sensor in flow generator*
Leak
Tidal volume
Respiratory rate
Minute ventilation
*
** The displayed values are estimates. They are provided for trending purposes only.
36
Range
-5 to 30 cm H
0–120 L/min
100–3000 mL
6–60 BPM
0.6–60 L/min
Results may be inaccurate in the presence of leaks.
Accuracy
O
±0.5 cm H
O
2
2
**
**
±0.5 BPM
**
Display
Resolution
0.1 cm H
O
2
1 L/min
1 mL
0.1 BPM
0.25 L/min

Hide quick links: