Important - Be Sure To Read This Chapter For Safe Use Of The Equipment; Safety Precautions - Olympus SZX7 Instructions Manual

Stereomicroscope system
Hide thumbs Also See for SZX7:
Table of Contents

Advertisement

IMPORTANT
The SZX7 stereomicroscope is provided with the ESD (Electro-Static Discharge) protection specifications. It is provided
with electroconductive coating in the external finish to reduce the surface resistance and grounding lead wires in their
standard stand or system to eliminate static electricity from them.
CAUTION

SAFETY PRECAUTIONS

1. After the equipment has been used in an observation of a specimen that is accompanied with a potential of infection,
clean the parts coming in contact with the specimen to prevent infection.
· Moving this microscope is accompanied with the risk of
dropping the specimen. Be sure to remove the specimen
before moving this product.
· Hold the bottom of the stand with one hand while holding
base column with the other hand to avoid tilting the micro-
scope.
· In case the specimen is damaged by erroneous operation,
promptly take the infection prevention measures.
2. The desktop should be a level surface with an inclination of less than 3°, and the load to the microscope should be less
than 7 kg (when the SZ2-ST standard stand is used).
The microscope may become unstable when certain intermediate attachment and/or photography unit are mounted on
it. Be careful so that the microscope does not topple down.
3. When the microscope is to be disposed of, follow your local industrial waste disposal regulations.
If you have any questions, please contact Olympus.
1
Getting Ready
1. A microscope is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe impact.
2. Do not use the microscope where it is subjected to direct sunlight, high temperature and humidity, dust or vibrations. (For
the operating conditions, see Chapter 6, "SPECIFICATIONS" on page 14.)
3. Do not turn the zoom adjustment knobs beyond the stopper positions. Otherwise, the internal mechanism may be
damaged.
Do not turn the left and right zoom adjustment knobs in the opposite directions, as this will result in a failure.
4. Only one intermediate attachment having thickness up to 60 mm can be mounted on the microscope. If two intermedi-
ate attachments or an intermediate attachment thicker than 60 mm is mounted, the image may be cut a little.
However, the coaxial reflected light illuminator (SZX-ILLC) is not counted in the number of intermediate attachment.
}When using multiple modules, they should be stacked in order of the SZX-ILLC, SZX-AS, SZX-RFA (SZX-RFL), SZX-SDO,
SZX-APT, SZX-BS, SZX-DA, SZX-FAD and SZX-EPA from the bottom to the top.
1
To maintain the ESD protection performance, always use the microscope in combination with the
modules described in this manual or options having the ESD protection specifications. Otherwise,
the grounding will not work properly.