INSTRUCTIONS
SZ61/SZ61-60
SZ61TR
SZ51/SZ51-60
ZOOM STEREO MICROSCOPES
This instruction manual is for the Olympus Zoom Stereo Microscopes. To ensure the safety,
Optical Microscope and Accessory
obtain optimum performance and to familiarize yourself fully with the use of this microscope, we
recommend that you study this manual thoroughly before operating the microscope.
Retain this instruction manual in an easily accessible place near the work desk for future reference.
For details of products included in the configuration of this system, see "1 MODULES AND
NOMENCLATURE" (P. 3) and "8 OPTIONAL MODULES" (P. 18).
Downloaded from
ManualsNet.com
search engine

Advertisement

Table of Contents
loading

Summary of Contents for Olympus SZ61

  • Page 1 SZ61TR SZ51/SZ51-60 ZOOM STEREO MICROSCOPES This instruction manual is for the Olympus Zoom Stereo Microscopes. To ensure the safety, Optical Microscope and Accessory obtain optimum performance and to familiarize yourself fully with the use of this microscope, we recommend that you study this manual thoroughly before operating the microscope.
  • Page 2 Downloaded from ManualsNet.com search engine...
  • Page 3: Table Of Contents

    SZ2 series CONTENTS IMPORTANT ............................... 1 1 MODULES AND NOMENCLATURE ..................3 2 SUMMARY OF OBSERVATION PROCEDURE ..............5 2-1 Preparation ................................5 2-2 Observation Procedure ..........................5 3 OPERATION ..............................6 3-1 Base ....................................6 Using the Stage Plate ................................6 Adjusting the Focus Adjustment Knob Tension ..........................6 3-2 Microscope Body ..............................
  • Page 4 Downloaded from ManualsNet.com search engine...
  • Page 5: Important

    SZ2 series IMPORTANT The SZ2 series of stereo microscopes are provided with ESD (Electro-Static Discharge) protection specifications. They are provided with electroconductive coating in their external finish to reduce the surface resistance and grounding lead wires in their standard base or frame to eliminate static electricity from them. To maintain the ESD protection performance, always use the microscope in combination with the NOTE modules described in this manual or options having the ESD protection specifications.
  • Page 6 Maintenance and Storage 1. To clean the lenses and other glass components, simply blow dirty away using a commercially available blower and wipe gently using a piece of cleaning paper (or clean gauze). If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available absolute alcohol.
  • Page 7: Modules And Nomenclature

    (ESD protection on black surface) SZX2-ILLTS* · SP-C Accessory mount holes Optional stage mounting holes M4 screw (x 2) Olympus classifies SZ61, SZ61-60, SZ61TR, SZ51 and SZ51-60 as an optical microscope and other units as optical microscope accessories. Downloaded from ManualsNet.com search engine...
  • Page 8 Observation Tube of SZ61TR C-mount (Page 11) C-mount clamping screw Confocality adjustment clamping screw (Page 11) (x 2) Light path selector lever (Page 10) Pulled out : Camera path 100% (Right eyepiece sleeve light path 0%) Pushed in : Right eyepiece sleeve light path 100% Downloaded from ManualsNet.com...
  • Page 9: Summary Of Observation Procedure

    SZ2 series SUMMARY OF OBSERVATION PROCEDURE 2-1 Preparation Ref. Page 1. Confirm the mounting, particularly that of the microscope body............................(Page 15 — 17) 2. Confirm that the eyepieces are positioned correctly with respect to the standard base...................(Page 16) 3.
  • Page 10: Operation

    OPERATION 3-1 Base Using the Stage Plate When reflected light illumination is used, the stage plate can be placed either the white or black surface facing up depending on which side makes the specimen easier to view. · To maintain the ESD performance, use the black surface of NOTE the SZ2-SPBW stage plate.
  • Page 11: Adjusting The Diopter (Zoom Confocality Adjustment)

    SZ2 series Adjusting the Diopter (Zoom Confocality Adjustment) The adjustment operation is variable whether the eyepieces use an eyepiece micrometer disks or not. Eyepieces without eyepiece micrometer disk Turn the diopter adjustment rings b of the left and right eyepieces to positions “0”.
  • Page 12: Using The High/Low Zoom Magnification Stopper

    Using the High/Low Zoom Magnification Stopper The magnification can be limited in the desired range by setting each stopper on the microscope body (also used as the magnification indices) and the stopper ring on each zoom adjustment knob. The right zoom adjustment knob is used to limit the higher end of the magnification, and the left knob is used to limit the lower end.
  • Page 13: Mounting The Eyepiece Micrometer Disk

    SZ2 series Mounting the Eyepiece Micrometer Disk Indication WHSZ10X-H/15X-H/20X-H engraved surface Rotate the disk-mounting ring a of the eyepiece counterclockwise to remove. Prepare an eyepiece micrometer disk b (24 mm dia. x 1.5 mm thick), remove dust and dirt from its surface, and fit it into the disk-mounting ring a so that the engraving on the micrometer disk faces downward.
  • Page 14: Using The Auxiliary Objectives

    Using the Auxiliary Objectives All of the auxiliary objectives are provided with the ESD protection performance. · The 0.3X and 0.4X auxiliary objectives cannot be used in NOTE combination with the standard base because their working Microscope body distances are too long. ·...
  • Page 15: Adjusting The Confocality Of The Camera (Sz61Tr)

    SZ2 series Adjusting the Confocality of the Camera (SZ61TR) Ensuring the confocality between the observation image and the camera monitor image makes it possible to maintain accurate focusing of the observation image or monitor image even when the light path is switched. Adjust the diopter of the eyepiece (see page 7) and bring the specimen into focus.
  • Page 16: Troubleshooting Guide

    Under certain conditions, performance of the microscope may be adversely affected by factors other than defects. If problems occur, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact Olympus. Problem...
  • Page 17: Specifications

    SZ2 series SPECIFICATIONS Item SZ61 SZ51 SZ51-60 SZ61-60 SZ61TR Magnification 0.67X to 4.5X 0.8X to 4X Zoom ratio Working distance 110 mm Tube tilting angle 45° 60° 45° 60° Interpupillary Microscope Left/right interlocked. distance body Adjustment range: 52 to 76 mm (using the WHSZ10X eyepieces)
  • Page 18: Optical Performance

    OPTICAL PERFORMANCE The following data shows only the typical magnifications of each zoom magnification. Eyepieces Working WHSZ 10X WHSZ 15X WHSZ 20X WHSZ 30X Zoom Distance FN 22 FN 16 FN 12.5 FN 7 Magnification (mm) Total Actual Field Total Actual Field Total Actual Field...
  • Page 19: Assembly

    SZ2 series ASSEMBLY 7-1 Assembly Diagram The diagram below shows the sequence of assembly of the various modules. The numbers indicate the order of assembly. Assembly steps enclosed in will be detailed on the subsequent pages. When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid NOTE scratching any parts or touching glass surfaces.
  • Page 20: Detailed Assembly Procedures

    7-2 Detailed Assembly Procedures Installing the Microscope Body · Loosen the body clamping knob a and insert the microscope body gently. · The body clamping knob a can be attached to any of the three positions b around the arm. After installation, attach the knob to a position so that the knob does not come in the way of the light source, etc.
  • Page 21 SZ2 series Mounting the Camera (SZ61TR only) Using the Allen wrench, loosen the C-mount clamping screw a and remove the C-mount seat b . Screw the C-mount seat b into the camera c . Place the C-mount seat in the original position and tighten the clamping screw a .
  • Page 22: Optional Modules

    OPTIONAL MODULES 8-1 BX Stage Adapter Type 1 SZX-STAD1 Introduction The SZX-STAD1 is an adapter for use in mounting the U-SRG or U-SRP rotary stage on the SZ2-ST standard base, SZX2- ST standard base, large base or SZX2 illumination base. The U-SRP is used in combination with the U-FMP mechanical stage to enable movement in the X- and Y-directions, which is convenient for framing in photomicrography or camera observation.
  • Page 23 SZ2 series Installation Use the tools shown in the table below. Tool Description Allen wrench (for M4 screw) Provided with SZX-STAD1 Hexagon socket screw (M4) Provided with SZX-STAD1 Place the notch b of SZX-STAD1 a on the back side of the base, and secure the screws (2 pcs.) using the Allen wrench to attach SZX- STAD1 a to a base.
  • Page 24: Stage Adapter Type 1 Szh-Stad1

    8-2 Stage Adapter Type 1 SZH-STAD1 The SZH-STAD1 is an adapter providing the similar function to the BX stage adapter type 1 (SZX-STAD1). The applicable stage is the BH2-SH horizontal knob stage. Assembly STAD mount for ILLT SZX2-STADM is necessary for using the SZX2-ILLTQ/ILLTS LED transmitted light illumination base.
  • Page 25: Gliding Stage Szh-Sg

    SZ2 series 8-3 Gliding Stage SZH-SG External View and Nomenclature Stage plate Finger hook Finger hook Gliding stage Gliding stage Seat Illumination field diameter : 40 mm Moving range diameter : 40 mm Installation · Be sure to clean the sliding surface if dirt or metallic power is attached on them. NOTE ·...
  • Page 26: Cup Stage Szh-Sc

    8-4 Cup Stage SZH-SC External View and Nomenclature The SZH-SC can be used only with reflected light illumination. NOTE Tube Finger hook Cup stage Cup stage Tilting limit angle : 30° Tilting limit angle : 30° Stage plate Specimen holder Specimen holder Specimen holder mount holes Seat...
  • Page 27 SZ2 series Operation Place a specimen on the stage plate, hold the cup stage by the edge, and tilt the cup stage slowly. If the specimen slips on the stage plate, hold the specimen with the provided specimen holder. To fix a container such as a petri dish, insert the provided tube into the specimen holder to fix the container by pinching.
  • Page 28 Manufactured by Shinjuku Monolith, 2-3-1 Nishi-Shinjuku, Shinjuku-ku, Tokyo 163-0914, Japan Distributed by AX6611 10 Issued in March, 2019 Downloaded from ManualsNet.com search engine...