Page 1
RESEARCH STEREOMICROSCOPE SYSTEM This instruction manual is for the Olympus SZX Research Stereomicroscope System. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this microscope, we recommend that you study this manual thoroughly before operating the microscope. Retain this instruction manual in an easily accessible place near the work desk for future reference.
IMPORTANT SAFETY PRECAUTIONS 1. To prevent the microscope body from turning over, its pivot angle must be limited to 30° as shown in Fig. 1. 2. Sufficient care is required during observation with a magnification objec- tive, because the objective´s long working distance can displace the microscope body to a higher position.
4. Only one intermediate attachment can be used. Using two attachments may obscure a part of the image. However, the vertical illuminator (SZX- ILLC) is not considered to be an attachment. }When more than one intermediate attachment is to be combined, they should be stacked according to the following order, from the bottom to the top: SZX-ILLC, SZX-AS, SZX-RFA (SZX-RFL), SZX-SDO, SZX-APT, SZX- BS, SZX-DA, SZX-FAD, then SZX-EPA .
ASSEMBLY 2-1 Assembly Diagram The diagram below shows how to assemble the various modules. The numbers indicate the order of assembly. # When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts or touching glass surfaces.
2-2 Detailed Assembly Procedure Installing the Pillar (Fig. 3) 1. Using the Allen wrench provided with the base, loosen the 2 clamping ² screws @ on the pillar support sleeve completely. 2. Hold the pillar ² with the white rubber cap at the top, and insert it into the pillar support sleeve until it reaches the bottom.
Page 8
Mounting the Microscope Body (Fig. 6) 1. Remove the cap @ on the focusing assembly by inserting a thin object into the notch. 2. Using the provided Allen screwdriver, loosen the dovetail mount clamp- ing screw inside the cap on the focusing assembly by rotating it by 2 or ³...
Page 9
Mounting the Eyepiece (Fig. 9) 1. Remove the eyepiece dust caps @ and loosen the eyepiece clamping screws ² completely. 2. Gently insert the cross-lined eyepiece (CROSS WHS10X) ³ (if this is not provided, use the WHS10X-H) into the right eyepiece sleeve until it stops. ³...
SUMMARY OF OBSERVATION PROCEDURE 4-1 Preparation 1. Check and tighten the connection of each component, especially the observation tube............(Page 4) 2. Adjust the position of the microscope body......................................(Page 5) 3. Adjust the tension of the coarse focus adjustment knob..............................(Page 8) 4.
USING THE CONTROLS 5-1 Base Using the Stage Plate When reflected light is used, the stage plate is usually placed with the white side facing up. In case the specimen is white or other bright colors, place the stage plate with the black side up to improve contrast. # When transmitted light is used, use the transparent glass stage plate (SP-C).
Page 13
Engagement and Disengagement of the (Fig. 14) Zooming Knob Click Stop Function }When the click stop knob is set to ON, the click stop function is engaged for each magnification indicated with the zooming knob. When the knob is set to OFF, the zoom magnification can be varied continuously and finely near the click groove.
Page 14
Magnification Indicator Ring (Fig. 16) }Every objective other than the 1X objective is provided with an observa- tion magnification* indicator ring. ² * This refers to the total magnification when the 10X eyepiece is used. }When the revolving nosepiece (SZX-2RE) is used, the total observation magnifications when two objectives are switched can be read directly by mounting these rings on the left and right zooming knobs.
5-3 Observation Tube Interpupillary Distance Adjustment (Fig. 20) # Be sure to hold the binocular assembly @ with both hands to make this adjustment. While looking through the eyepieces, hold the left and right of the bin- ocular assembly @ and adjust the eyepieces by opening or closing them for binocular vision until the left and right fields of view coincide completely.
Page 16
Using the Eye Shades (Fig. 22) When Wearing Eyeglasses Use with eye shades in their normal folded-down position. This will pre- vent eyeglasses from being scratched by the eyepiece. When Not Wearing Eyeglasses Extend the folded eyeshades in the direction of the arrow. This makes observation easier by preventing the inverse incidence of light from be- tween the eyepiece and your eyes.
Page 17
Light Path Selection (SZX-TR30) (Fig. 24) Slide the light path selector knob @ to select the desired light path. Light Path Indication Intensity Ratio Selector Knob Pushed in 100% for binocular eyepieces. Pulled out 20% for binocular eyepieces. 80% for photomicrography and TV observation.
5-4 TV Observation and Photomicrography }When TV observation or photomicrography is required, use the SZX-TR30 trinocular tube. A TV camera and/or digital camera unit can be mounted on the SZX-TR30 trinocular tube by means of the TV adapter and/or camera mount adapter*. * The camera mount adapter is not necessary if a TV adapter equipped with a camera mount is used.
Page 20
Item Specification 6. Objectives SZX-ZB12 SZX-ZB9 WD: Working distance Model Model PF: Parfocal lens DFPLFL0.3X 130 mm* DFPL0.5X-4 171 mm* DFPLFL0.45X 198 mm* DFPL0.75X-4 116 mm DFPLFL0.5XPF 70 mm DFPLAPO1X-4 81 mm DFPLAPO1XPF 74 mm SZX-ACH1X 90 mm DFPLAPO1.2XPF2 60 mm SZX-ACH1.25X 68 mm * Auxiliary pillar is required when...
OPTICAL CHARACTERISTICS ■ SZX-ZB12 Eyepiece WHS10X-H WHS15X-H WHS20X-H WHS30X-H Objective Field of view Field of view Field of view Field of view Total Mag. Total Mag. Total Mag. Total Mag. (mm) (mm) (mm) (mm) DFPLFL0.3X 2.1X - 27X 104.8 - 8.1 3.15X - 40.5X 76.2 - 5.9 4.2X - 54X 59.5 - 4.6...
Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a problem occurs, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact your local Olympus representative for assistance. Problem...
OPERATION OF OTHER ATTACHMENTS 9-1 Eyepoint Adjuster SZX-EPA External View Observation tube clamping screw Assembly (Fig. 27) 1. Using the Allen screwdriver provided with the SZX microscope body, remove the observation tube @. 2. Mount the eyepoint adjuster ² to the place where the observation tube ²...
9-2 AS Unit SZX-AS (for use with the SZX-ZB9) External View Aperture iris diaphragm adjustment Observation tube clamping screw lever Assembly }The AS unit can be mounted in the same manner as the SZX-EPA eyepoint adjuster, which is described in section 9-1. Using the Aperture Iris Diaphragm }By adjusting the aperture iris diaphragm, it is possible to improve the contrast of the observed image and increase the depth of focus.
9-3 Drop Prevention Collar SZX-R and Auxiliary Pillar SZH-P400/SZH-P600 }The auxiliary pillar is to be used when observing a large specimen or when it is required to move the microscope body height using a low-magnification objective. }The drop prevention collar prevents the zoom microscope body from dropping when installed at a high position using the auxiliary pillar and the clamping knob on the focusing assembly is loosened carelessly.
9-4 Revolving Nosepiece SZX-2RE External View Hand guard Allen wrench (for M3 screws) M3 –– 4 mm long (x 2) Objective mounts Clamping Nosepiece mount screws M3 –– 6 mm long (x 3) Assembly (Figs. 29 - 32) 1. Attach the hand guard ² to the revoling noseplece @ by tightening the ³...
Page 27
5. Place the nosepiece mount … (with its objective mounts † facing … † ‡ upward) where the objective has been by aligning the screw holes. Š Using the Allen wrench (for M3 screws), clamp the revolver mount with the provided 3 clamping screws (M3 –– 6 mm long). (Fig. 31) # As the screw holes may be hidden behind the objective mounts †, clamp the nosepiece mount while rotating it.
9-5 BX Stage Adapter Type 1 SZX-STAD1 Introduction This adapter is for installation of a U-SRG or U-SRP rotary stage on the SZX standard base or a SZX series illumination base. When the U-SRP rotary stage is used together with the U-FMP mechanical stage, X-Y directional movement be- comes possible, which is convenient for framing during photomicrography.
Page 29
Assembly Mechanical stage U-FMP Rotary stage Rotary stage U-SRP U-SRG Clamping screws Allen screwdriver (provided with the microscope body) BX stage adapter type 1 SZX-STAD1 Clamping screw Attaching hole Attaching threaded hole Base Mounting the Polarizer (SZX-PO) When simplified transmitted polarized light observation is required, install the polarizer on the SZX-STAD1 BX stage adapter type 1.
9-6 BX Stage Adapter Type 2 SZX-STAD2 Introduction This adapter is for installing the U-SIC large stage* on the SZX-STL large stand**. When this stage adapter is used, the SZH-P400 auxiliary pillar should be used to cover the height of the stage adapter. In addition, when an objective with a long working distance (DFPLFL0.3X, DFPLFL0.45X or DFPL0.5X) is used, the SZH-P600 auxiliary pillar should be used (always together with the drop prevention collar).
Simplified Transmitted Light Observation (Fig. 34) 1. Illuminate the specimen with an external light source (LSD illuminator, Microscope light guide illuminator, etc.). body Light the external light source as shown in the figure on the left and Objective Frosted filter irradiate the mirror assembly.
Page 34
2-43-2, Hatagaya, Shibuya-ku, Tokyo, Japan Postfach 10 49 08, 20034, Hamburg, Germany 2 Corporate Center Drive, Melville, NY 11747-3157, U.S.A. One Corporate Drive, Orangeburg, NY 10962, U.S.A. 491B River Valley Road, #12-01/04 Valley Point Office Tower, Singapore 248373 2-8 Honduras Street, London EC1Y OTX, United Kingdom. 31 Gilby Road, Mt.