Troubleshooting; Troubleshooting Table - Invacare 1556861 User Manual

Ceiling hoist
Table of Contents

Advertisement

I-Transia™

7 Troubleshooting

7.1 Troubleshooting Table

SYMPTOMS
I-Transia hoist does not
respond to the hand control's
keys.
Trolley does not traverse.
Hoist will not charge
No indicator light on hoist
The emergency lowering cord
does not turn hoist off
The emergency lowering cord
does not lower the hoist.
If problems are not remedied by the suggested means, please contact your
Invacare dealer.
22
PROBLEM
Emergency stop is activated.
System does not have power
or battery is discharged
Rails require service or
cleaning.
System does not have power
System does not have power
or battery is discharged.
Hoist requires service.
Hoist requires service.
SOLUTION
Check that the emergency
stop is not activated.
Check that the I-Transia has
power supply and that the
battery is recharged.
Check that the transformer
is switched on and connected
to the charging station.
Place the hand control in the
charging station and recharge
the I-Transia.
Contact an Invacare dealer
for service.
Place the hand control in the
charging station and recharge
the I-Transia
Check that the power supply
is switched on and connected
to the charging station.
Contact an Invacare dealer
for service.
Check that the I-Transia has
power supply and that the
battery is recharged.
Check that the power supply
is switched on and connected
to the charging station.
Place the hand control in the
charging station and recharge
the I-Transia.
Contact an Invacare dealer
for service.
Contact an Invacare dealer
for service.
1183377-A