Making Measurements

Making Measurements
The Agilent N8900 series has fully integrated voltmeter and ammeter to measure the actual voltage and current that
is being supplied to the load. Whenever the instrument is on, the front panel automatically measures output voltage
and current.
Power measurements are derived from the voltage and current measurements.
Average (DC) Measurements
Use the following commands to make a measurement:
Front Panel
Select the Meter key.
Repeatedly press this key to cycle
through the following measurement
functions:
Voltage, Current
Voltage, Power
Voltage, Current, Power
58
SCPI Command
To measure average (DC) output
voltage, current, or power:
MEAS:VOLT?
MEAS:CURR?
MEAS:POW?