J207 Voltage Measurements

Cable W7 connects the A1 Main Board Assembly J307 to the A2 Interface Assembly J207. Table 3-4
provides a listing of the voltages between these assemblies. Flow charts in Figure 3-1 will refer to several of
these voltages to determine if they are outside the normal range.
Table 3-4. Voltage Measurements at J207 (A2 Interface to A1 Main board)
A2J207
Signal Name
Pin #
1
PM_INHIBIT (Enabled)
2
OV_SCR*
3
OV_PROG
4
FAN_PROG
5
OV_DETECT*
6
SW_POS (Norm)
7
RANGE_SELECT (High)
8
OS_TRIM_NEG (COMP)
OS_TRIM_NEG (SCPI)
9
+5Vs
10
COMMON
11
COMMON
12
+15Vs
13
-15Vs
14
HS_THERM (@25C)
15
FUSE
16
IMON_H
17
IMON_L
IMON_L (@20mA Out)
18
IMON_P
19
VMON
20
COMMON
21
COMMON
22
COMMON
23
COMMON
24
CV_PROG
25
CC_PROG
26
CC_DETECT*
27
CCN_DETECT*
28
CV_DETECT*
CV Mode
Full Scale Voltage
Full Scale Voltage
No Load
0
+5
+3.9
+2.8
+5
+5
0
+1.7
+4.0
+5
0
0
+15
-15
+2.5
+2.4
0
0
+4.8
0
+4.8
0
0
0
0
-4.8
-4.8
+5
+5
0
Troubleshooting - 3
CC Mode
Signal Origin
Full Load
0
+5
+3.9
+3.8
+5
+5
0
+1.7
+4.0
+5
0
0
+15
-15
+2.5
+2.6
+3.5
+14.7
+4.8
0
+4.8
0
0
0
0
-4.8
-4.8
0
+5
+5
A2
A2
A2
A2
A1
A1
A2
A2
A2
A1
A1
A1
A1
A1
A1
A1
A1
A1
A1
A1
A1
A1
A1
A2
A2
A1
A1
A1