Invacare FDX User Manual page 93

Power wheelchair base
Hide thumbs Also See for FDX:
Table of Contents

Advertisement

Service Indicator Light Diagnostics
For this procedure, refer to FIGURE 1.
NUMBER OF
ERROR CODE
FLASHES
DESCRIPTION
1
User Fault
2
Battery Fault
3
Left Motor Fault
4
Right Motor Fault
5
Left Park Brake Fault
6
Right Park Brake Fault
Part No 1163181
FIGURE 1 Service Indicator Light Diagnostics
Release joystick to neutral and try again.
Charge the batteries. Refer to Charging Batteries on page 53.
Check that battery cables are connected properly. If necessary, replace batteries. Refer to
Replacing the Batteries and/or Battery Cables on page 83.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is
plugged into the controller.
Contact Invacare/Dealer for service.
Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is
plugged into the controller.
Contact Invacare/Dealer for service.
93
Service Indicator Light
POSSIBLE SOLUTION
9 TROUBLESHOOTING
Invacare® FDX®

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fdx-cgFdx-mcg

Table of Contents